Table 3-6. Analog Output Zero To Full Scale Values - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Choose Signal To Output
Thermo Fisher Scientific
In the Main Menu, choose Instrument Controls > IO Configuration >
Analog Output Config > Select Channel > Set Minimum or Maximum
Value.
MINIMUM OUTPUT PERCENT:
SELECTED OUTPUT:
CURRENTLY:
SET TO:
CHANGE VALUE
RANGE
Table 3–6. Analog Output Zero to Full Scale Values
Output
Zero % Value
SO2
Zero (0)
LO SO2
Zero (0)
HI SO2
Zero (0)
Range SO2
Recommend not to change the setting for this output
Internal Temp
User-set alarm min value
Chamber Temp
User-set alarm min value
Perm Oven Gas Temp
Perm Oven Gas alarm min value
Perm Oven Heater
Perm Oven Heater alarm min
Temp
value
Pressure
User-set alarm min value
Sample Flow
User-set alarm min value
PMT Volts
700 volts
Flash Voltage
User-set alarm min value
Flash Reference
User-set alarm min value
Everything Else
0 units
The Choose Signal To Output screen displays a submenu list of the analog
output signal group choices. Group choices are Concentrations, Other
Measurements, and Analog Inputs (if the I/O expansion board is installed).
This allows the user to select the output signal to the selected output
channel. In dual or auto range mode, "HI" or "LO" is displayed to indicate
high or low concentrations. Range (SO
The Concentrations screen is shown below. See Table 3-8 that follows for a
list of items for each signal group choice.
V 1
N/A %
0000.5
SAVE
AVG
DIAGS ALARM
) is visible only in auto range mode.
2
Model 43i Trace Level-Enhanced Instruction Manual
Operation
Instrument Controls Menu
% ?
Full-Scale 100% Value
Range Setting
Range Setting
Range Setting
User-set alarm max value
User-set alarm max value
Perm Oven Gas alarm max value
Perm Oven Heater alarm max
value
User-set alarm max value
User-set alarm max value
1100 volts
User-set alarm max value
User-set alarm max value
10 units
3-43