Operation
Service Menu
ANALOG INPUT CAL:
>INPUT CHANNEL 1
INPUT CHANNEL 2
INPUT CHANNEL 3
INPUT CHANNEL 4
INPUT CHANNEL 5
INPUT CHANNEL 6
INPUT CHANNEL 7
RANGE
INPUT CHANNEL 8
Analog Input Calibration
Analog Input Calibrate
3-74
Model 43i Trace Level-Enhanced Instruction Manual
AVG
DIAGS ALARM
The Analog Input Calibrate Zero screen allows the user to calibrate the
Zero
zero state of the selected analog input.
In the Main Menu, choose Service > Analog Input Calibration > Select
Channel > Calibrate Zero. (Hook up a voltage source of 0 V to the analog
input channel.)
The Analog Input Calibration Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog input.
Full-Scale
In the Main Menu, choose Service > Analog Input Calibration > Select
Channel > Calibrate Full Scale. (Hook up a voltage source of 10 V to the
analog input channel.)
ANALOG INPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE
ANALOG INPUT CAL:
DISCONNECT SELECTED INPUT!
SELECTED INPUT:
CURRENTLY:
CALIBRATE INPUT TO ZERO
RANGE
AVG
ANALOG INPUT CAL:
PROVIDE VOLTAGE TO INPUT!
SELECTED INPUT:
CURRENTLY:
SET TO:
CALIBRATE TO VALUE
RANGE
AVG
AVG
DIAGS ALARM
ZERO
INPUT 1
0.00 V ?
DIAGS ALARM
SPAN
INPUT 1
9.80 V
10.00 V ?