Analog Input Calibration; Calibrating The Input Channels To Zero Volts - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Servicing

Analog Input Calibration

Analog Input
Calibration
Calibrating the Input
Channels to Zero
7-24
Model 43i Trace Level-Enhanced Instruction Manual
The Analog Output Cal line displays Zero.
Note When calibrating the analog output, always calibrate zero first and
then calibrate full scale.
5. Use
Output To line, then press
6. Press
7. Press
8. Use
Output To line, then press
Use the following procedures to calibrate the analog inputs after replacing
the optional I/O expansion board. These procedures include selecting
analog input channels, calibrating them to zero volts, and then calibrating
them to full scale using a known voltage source.
Use the following procedure to calibrate the input channels to zero volts.
1. From the Main Menu, press
to scroll to Analog Input Calibration, and press
The Analog Input Cal screen displays.
Volts
Note If Service Mode is not displayed, refer to "Accessing the Service
Mode" in this chapter, then return to the beginning of this step.
2. At the Analog Input Cal screen, press
press
3. With the cursor at Calibrate Zero, press
The screen displays the input voltage for the selected channel.
until the meter reads the value shown in the Set
to return to the previous screen.
to select Calibrate Full Scale.
until the meter reads the value shown in the Set
.
to save the value.
to save the value.
to scroll to Service, press
to scroll to a channel, and
.