Calibrate Pressure Zero; Calibrate Pressure Span - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Operation
Service Menu

Calibrate Pressure Zero

Calibrate Pressure Span

3-68
Model 43i Trace Level-Enhanced Instruction Manual
The Calibrate Pressure Zero screen calibrates the pressure sensor at zero
pressure.
Note A vacuum pump must be connected to the pressure sensor before
performing the zero calibration.
In the Main Menu, choose Service > Pressure Calibration > Zero.
CALIBRATE PRESSURE ZERO:
CURRENTLY:
SET TO:
CONNECT VACUUM PUMP AND
SAVE ZERO PRESSURE
RANGE
The Calibrate Pressure Span screen allows the user to view and set the
pressure sensor calibration span point.
Note The plumbing going to the pressure sensor should be disconnected so
the sensor is reading ambient pressure before performing the span
calibration. The operator should use an independent barometer to measure
the ambient pressure and enter the value on this screen before
calibrating.
In the Main Menu, choose Service > Pressure Calibration > Span.
CALIBRATE PRESSURE SPAN:
CURRENTLY:
SET TO:
CHANGE VALUE
RANGE
0.0 mmHg
0.0 mmHg ?
AVG
DIAGS ALARM
756.0 mmHg
760.0 mmHg ?
MOVE CURSOR
SAVE
AVG
DIAGS ALARM