Calibrate Flow Zero; Calibrate Flow Span; Restore Default Flow Calibration - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Operation
Service Menu

Calibrate Flow Span

Restore Default Flow
RESTORE DEFAULT CAL:
RANGE
3-70
Model 43i Trace Level-Enhanced Instruction Manual
In the Main Menu, choose Service > Flow Calibration > Zero.
The Calibrate Flow Span screen allows the user to view and set the flow
sensor calibrate span point.
Note An independent flow sensor is required to read the flow, then the
operator enters the flow value on this screen to perform the calibration.
In the Main Menu, choose Service > Flow Calibration > Span.
The Restore Default Flow Calibration screen allows the user to reset the
configuration values to factory defaults.
Calibration
In the Main Menu, choose Service > Flow Calibration > Set Defaults.
AVG
DIAGS ALARM

CALIBRATE FLOW ZERO:

CURRENTLY:
SET TO:
DISCONNECT PUMP AND
SAVE ZERO PRESSURE
RANGE
AVG
CALIBRATE FLOW SPAN:
CURRENTLY:
SET TO:
CHANGE VALUE
RANGE
AVG
RESTORE DEFAULT CAL:
RESTORE
ARE YOU SURE YOU WANT
PRESS
RANGE
0.000 LPM
0.000 LPM ?
DIAGS ALARM
0.508 LPM
0.808 LPM ?
MOVE CURSOR
SAVE
DIAGS ALARM
TO CONFIRM RESTORE
AVG
DIAGS ALARM
RESTORE
TO?x