Measurements Reported In Response To Da Command; Single Range Mode; Dual/Auto Range Mode - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Geysitech (Bayern-Hessen) Protocol
Geysitech Commands
Measurements reported in
response to DA command

Single Range Mode

Dual/Auto Range Mode

D-6
Model 43i Trace Level-Enhanced Instruction Manual
Example:
Geysitech Protocol with transmission of three concentrations (Instrument
ID is 1, Operation Status is 03, Error Status is 04):
Data Query String:
Reply String:
<STX>MD03<SP>001<SP>+2578+01<SP>03 <SP>04<SP>0000000000 <SP>002 <SP>
Address First Concentration(E-format)=25.78
+5681+00<SP>03<SP>04<SP>0000000000<SP>003<SP>+1175+01<SP>03<SP>04<SP
Second Concentration = 5.681
0000000000<SP><CR>
The attached concentrations are in the selected gas units. The
measurements that are attached to the response if not valid in a particular
mode then a value of 0.0 will be reported.
The following measurements reported in response to DA command are for
the Model 43i Trace Level-Enhanced.
The 1 measurement reported in single range mode includes:
SO
2
The 2 measurements reported in dual or auto range modes include:
low SO
2
high SO
2
<STX>DA<CR>
Address+2
Address+1
Third Concentration=11.75