Download Print this page

Number Of Records; Date And Time - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Operation
Instrument Controls Menu

Number of Records

Date and Time

3-24
Model 43i Trace Level-Enhanced Instruction Manual
The Number of Records screen is used to select the starting record to
display the number of records back to view.
SET # BACK FROM CURRENT:
TOTAL LRECS:
CHANGE VALUE
RANGE
AVG
The Record Display screen (read only) displays the selected records.
time
date
10:01 06/20/05 FC0088900
10:02 06/20/05 FC0088900
10:03 06/20/05 FC0088900
10:04 06/20/05 FC0088900
PGUP/DN
RANGE
AVG
The Date and Time screen is used to set a start date and time for which to
view logged data. For example, if "20 JUN 2005 10:00" is entered, then
the first logged data record that is displayed is the first record after this
time. If set to one minute logging, this would be at "20 JUN 2005 10:01".
DATE AND TIME:
20 JUN 2005
CHG
SET CURSOR TO MONTHS
ACCEPT AS SHOWN
RANGE
AVG
0000000
20
MOVE CURSOR
SAVE
DIAGS ALARM
flags
PAN L/R
DIAGS ALARM
10:00
DAYS
DIAGS ALARM
Thermo Fisher Scientific