Download Print this page

Communications Configuration - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

C-Link Protocol Commands

Communications Configuration

Communications
Configuration
B-34
Model 43i Trace Level-Enhanced Instruction Manual
Send:
set pmt off
Receive:
set pmt off ok
save
set save params
This command stores all current parameters in FLASH memory. It is
important that each time instrument parameters are changed, that this
command be sent. If changes are not saved, they will be lost in the event of
a power failure. The example that follows saves the parameters to FLASH
memory.
Send:
set save params
Receive:
set save params ok
time
This command reports the current time (24-hour time). The example that
follows reports that the internal time is 2:15:30 pm.
Send:
time
Receive:
time 14:15:30
set time hh:mm:ss
hh = hours
mm = minutes
ss = seconds
This command sets the internal clock (24-hour time). The example that
follows sets the internal time to 2:15 pm.
Note If seconds are omitted, the seconds default to 00.
Send:
set time 14:15
Receive:
set time 14:15 ok
addr dns
This command reports the TCP/IP address for the domain name server.
Send:
addr dns
Receive:
addr dns 192.168.1.1
set addr dns address
This command sets the dns address, where address consists of four numbers
ranging from 0-255 inclusive, separated by ".".
Thermo Fisher Scientific

Advertisement

loading