Thermo Scientific 43i Instruction Manual page 260

Pulsed fluorescence so2 analyzer

Advertisement

C-Link Protocol Commands
Datalogging
B-20
Model 43i Trace Level-Enhanced Instruction Manual
lrec per
srec per
These commands report the long and short records logging period. The
example that follows shows that the short record logging period is 5
minutes.
Send:
srec per
Receive:
srec per 5 min
set srec per value
set srec per value
value = | 1 | 5 | 15 | 30 | 6 0 |
These commands set the long and short records logging period to value in
minutes. The example that follows sets the long record logging period to 15
minutes.
Send:
set lrec per 15
Receive:
set lrec per 15 ok
no of lrec
no of srec
These commands report the number of long and short records stored in the
long and short records memory. The example that follows shows that 50
long records have been stored in the memory.
Send:
no of lrec
Receive:
no of lrec 50 recs
malloc lrec
malloc srec
These commands report the currently set memory allocation for long and
short records in percent of total memory.
Send:
malloc lrec
Receive:
malloc lrec 10%
set malloc lrec value
set malloc srec value
value = 0 to 100
These commands set the percent of memory space allocated for long and
short records to value, where value is a floating-point number representing
Thermo Fisher Scientific

Advertisement

loading