Instrument Action; Analog Output Configuration - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Instrument Action

Analog Output
Configuration
Thermo Fisher Scientific
The Logic State screen is used to change the I/O relay to either normally
Logic State
open or normally closed. The default state is open, which indicates that a
relay connected between the digital input pin and ground is normally open
and closes to trigger the digital input action. If nothing is connected to the
digital input pin, the state should be left at open to prevent the action from
being triggered.
Press
The Instrument Action screen allows the user to select the instrument state
that is tied to the selected digital input.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Digital Input Settings > Select Relay > Instrument State.
The Analog Output Configuration menu displays a list of the analog
output channels available for configuration. Channel choices include all
voltage channels, all current channels, voltage channels 1-6, and current
channels 1-6 (if the I/O expansion board option is installed). Configuration
choices include selecting range, setting minimum/maximum values, and
choosing signal to output.
to toggle and set the logic state open or closed.
DIGITAL INPUT SETUP:
>LOGIC STATE
INSTRUMENT ACTION
RANGE
AVG
CHOOSE ACTION:
>NONE
ZERO MODE
SPAN MODE
SET BACKGROUND
CAL TO LO SPAN
AOUTS TO ZERO
RANGE
AVG
AOUTS TO FS
CAL TO HI SPAN
Model 43i Trace Level-Enhanced Instruction Manual
OPEN
DIAGS ALARM
<--
DIAGS ALARM
Operation
Instrument Controls Menu
3-41