C-Link Protocol Commands
I/O Configuration
B-42
Model 43i Trace Level-Enhanced Instruction Manual
list var aout
list var dout
list var din
These commands report the list of index numbers, and the variables
(associated with that index number) available for selection in the current
mode (determined by single/dual/auto, gas mode) for analog output, digital
output and digital inputs. The index number is used to insert the variable
in a field location in a list using "set sp field index". The example that
follows reports the list of analog output, index numbers, and variables.
Send:
list var aout
Receive:
list var aout
index variable
0 none
1 so2
11 intt
12 rctt
16 pres
17 smplfl
18 pmtv
19 lmpv
20 lmpi
21 ain1
22 ain2
23 ain3
24 ain4
25 ain5
26 ain6
27 ain7
28 ain8
relay stat
This command reports the current relay logic normally "open" or normally
"closed," if all the relays are set to same state, that is all open or all closed.
The example that follows shows that the status when all the relays logic is
set to normally "open".
Send:
relay stat
Receive:
relay stat open
Note If individual relays have been assigned different logic then the
response would be a 4-digit hexadecimal string with the least significant
byte (LSB) being relay no 1.