C-Link Protocol Commands
Measurement Configuration
B-30
Model 43i Trace Level-Enhanced Instruction Manual
Send:
gas mode
Receive:
gas mode sample
set sample
set sample gas
This command sets the zero/span valves to the sample mode. The example
that follows sets the instrument to sample mode, that is, the instrument is
reading the sample gas.
Send:
set sample
Receive:
set sample ok
set zero
set zero gas
This command sets the zero/span valves to the zero mode. The example
that follows sets the instrument to zero mode that is, the instrument is
reading the sample gas.
Send:
set zero
Receive:
set zero ok
set span
set span gas
This command sets the zero/span valves to the span mode. The example
that follows sets the instrument to span mode that is, the instrument is
sampling span gas.
Send:
set span
Receive:
set span ok
gas unit
This command reports the current gas units (ppb or μ g/m
reports that the gas unit is set to ppm.
Send:
gas unit
Receive:
gas unit ppm
set gas unit
3
unit = | ppm | mg/m
|
This command sets the gas units to ppm or mg/m
follows sets the gas units to mg/m
Send:
set gas unit mg/m3
Receive:
set gas unit mg/m3 ok
3
. The example that
3
.
3
). The example