percent. The example that follows sets the memory allocation for long
records to 10.
Note Issuing these commands will clear all the logging data memory. All
the existing records should be retrieved using appropriate commands, if
required.
Send:
set malloc lrec 10
Receive:
set malloc lrec 10 ok
data treatment lrec
data treatment srec
These commands report the current selection of data treatment for
concentrations in the long records (lrecs) or short records (srecs). The
example that follows reports the data treatment for concentrations in lrec is
minimum.
Send:
data treatment lrec
Receive:
data treatment lrec min
set data treatment lrec string
set data treatment srec string
string = | cur | avg | min | max |
These commands set the data treatment to string, where string is current,
average, minimum, or maximum for the concentration values recorded in
the long records (lrecs) or short records (srecs). The example that follows
sets the data treatment for concentrations in lrec to minimum.
Send:
set data treatment lrec min
Receive:
set data treatment lrec min ok
set copy sp to lrec
set copy sp to srec
set copy sp to stream
These commands copy the current selections in scratch pad (sp) into the
long record, short record, or streaming data list. The example that follows
copies the current list in scratch pad into the long records list.
Send:
set copy sp to lrec
Receive:
set copy sp to lrec ok
C-Link Protocol Commands
Model 43i Trace Level-Enhanced Instruction Manual
Datalogging
B-21