I/O Configuration - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

C-Link Protocol Commands

I/O Configuration

I/O Configuration
B-38
Model 43i Trace Level-Enhanced Instruction Manual
set instrument id value
This command sets the instrument id to value, where value is a decimal
number between 0 and 127 inclusive.
Note Sending this command via RS-232 or RS-485 will require the host to
use the new id for subsequent commands.
Send:
set instrument id 12
Receive:
set instrument id 12 ok
mode
This command reports what operating mode the instrument is in: local,
service, or remote. The example that follows shows that the instrument is
in the remote mode.
Send:
mode
Receive:
mode remote
set mode local
set mode remote
These commands set the instrument to local or remote mode. The example
that follows sets the instrument to the local mode.
Send:
set mode local
Receive:
set mode local ok
program no
This command reports the analyzer's model information and program
version number, which will be dependant on the current version.
Send:
program no
Receive:
program no iSeries 43i 01.01.10.003
set layout ack
This command disables the stale layout/layout change indicator ('*') that is
attached to each response if the layout has changed.
Send:
set layout ack
Receive:
set layout ack ok
analog iout range channel
This command reports the analog current output range setting for channels,
where channel must be between 1 and 6, inclusive. The example that
follows reports current output channel 4 to the 4-20 mA range, according