C-Link Protocol Commands
Communications Configuration
B-36
Model 43i Trace Level-Enhanced Instruction Manual
baud
This command reports the current baud rate for the serial port
(RS232/RS485). The example that follows reports that the current baud
rate is 9600 baud.
Send:
baud
Receive:
baud 9600
set baud rate
rate = | 1200 | 2400 | 4800 | 9600 | 19200 | 38400 | 57600 | 115200 |
This command sets the instrument baud rate. The example that follows sets
the instrument's baud rate to 9600.
Note After the command is sent, the baud rate of the sending device must
be changed to agree with the instrument.
Send:
set baud 9600
Receive:
set baud 9600 ok
dhcp
This command reports the current state of use of DHCP on or off. DHCP
is used to assign an IP address to the analyzer automatically. The example
that follows shows that DHCP is on.
Send:
dhcp
Receive:
dhcp on
set dhcp onoff
These commands enables and disables the DHCP service by either on or
off. Changes to this parameter will only take effect when the analyzer is
powered up. The example that follows sets the DHCP service on.
Note When DHCP is set to on, the user-supplied addr gw, addr dns, addr
ip, and addr nm parameters are not used.
Send:
set dhcp on
Receive:
set dhcp on ok
format
This command reports the current reply termination format. The example
that follows shows that the reply format is 00, which means reply with no
checksum, according to Table B–9.
Thermo Fisher Scientific