Sample Flow

Lamp Intensity

Optical Span Test

Thermo Fisher Scientific
The Sample Flow screen (read only) displays the current sample flow rate.
The flows is measured by an internal flow sensor.
In the Main Menu, choose Diagnostics > Flow.
SAMPLE FLOW:
RANGE
The Lamp Intensity screen (read only) displays the current lamp intensity
in Hertz. Normally the intensity would read 90%, but this number will
decrease over time as the lamp degrades.
In the Main Menu, choose Diagnostics > Lamp Intensity.
LAMP INTENSITY:
RANGE
The Optical Span Test screen is used to turn the optical span test LED on
or off, and displays the SO2 concentration reading. Within the
fluorescence chamber is a light emitting diode (LED) which may be used to
simulate a particular concentration of SO
and easy way of checking the optics and electronics for span drift or other
problems.
In the Main Menu, choose Diagnostics > Optical Span Test.
OPTICAL SPAN TEST:
CURRENTLY:
SET TO:
RANGE
0.500 LPM
AVG
DIAGS ALARM
AVG
DIAGS ALARM
. It is designed to provide a quick
2
SO2:
TOGGLE VALUE
AVG
DIAGS ALARM
Model 43i Trace Level-Enhanced Instruction Manual
Operation
Diagnostics Menu
90 %
100
ON
OFF ?