Temperature
Min and Max Chamber
Temperature Limits
Thermo Fisher Scientific
Chamber
The Chamber Temperature screen displays the current chamber
temperature and sets the minimum and maximum alarm limits. The
chamber temperature is regulated at 45.0 degrees C. Acceptable alarm
limits range from 43 to 47 °C. If the chamber temperature reading goes
beyond either the minimum or maximum limit, an alarm is activated. The
word alarm appears in the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > Chamber Temp.
The Minimum Chamber Temperature alarm limit screen is used to change
the minimum chamber temperature alarm limit. The minimum and
maximum internal temperature screens function the same way.
In the Main Menu, choose Alarms > Chamber Temp > Min or Max.
Pressure
The Pressure screen displays the current reaction chamber pressure reading
and sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 400 to 1000 mmHg. If the pressure reading goes beyond either
INTERNAL TEMPERATURE:
ACTUAL MIN:
SET MIN TO:
INC/DEC
SAVE VALUE
RANGE
AVG

CHAMBER TEMPERATURE:

ACTUAL
>MIN
MAX
RANGE
AVG
CHAMBER TEMPERATURE:
ACTUAL MIN:
SET MIN TO:
INC/DEC
SAVE VALUE
RANGE
AVG
Model 43i Trace Level-Enhanced Instruction Manual
o
15.0
C
o
10.0
C ?
DIAGS ALARM
o
45.4
C
o
43.0
C
o
47.0
C
DIAGS ALARM
o
45.4
C
o
44.0
C
?
DIAGS ALARM
Operation
Alarms Menu
3-59