C-Link Protocol Commands
Alarms
B-12
Model 43i Trace Level-Enhanced Instruction Manual
alarm converter temp min
alarm converter temp max
These commands report the converter temperature alarm minimum and
maximum value current settings. The example that follows reports that the
converter temperature alarm minimum value is 205.0 °C.
Send:
alarm converter temp min
Receive:
alarm converter temp min 205.0 deg C
set alarm converter temp min value
set alarm converter temp max value
These commands set the converter temperature alarm minimum and
maximum values to value, where value is a floating-point number
representing converter temperature alarm limits in degrees C. The example
that follows sets the converter temperature alarm maximum value to
215.0 °C.
Send:
set alarm converter temp max 215
Receive:
set alarm converter temp max 215 ok
alarm internal temp min
alarm internal temp max
These commands report the internal temperature alarm minimum and
maximum value current settings. The example that follows reports that the
internal temperature alarm minimum value is 15.0 °C.
Send:
internal temp alarm min
Receive:
internal temp alarm min 15.0 deg C
set internal temp alarm min value
set internal temp alarm max value
These commands set the internal temperature alarm minimum and
maximum values to value, where value is a floating-point number
representing internal temperature alarm limits in degrees C. The example
that follows sets the internal temperature alarm maximum value to
45.0 °C.
Send:
set internal temp alarm max 45
Receive:
set internal temp alarm max 45 ok