Calibration Menu

Period Hours

3-20
Model 43i Trace Level-Enhanced Instruction Manual

Next Time

The Next Time screen is used to set the initial date and time of the
zero/span check. Once the initial zero/span check is performed, the date
and of the next zero/span check is calculated and displayed.
In the Main Menu, choose Calibration > Zero/Span Check > Next Time.
NEXT DATE AND TIME:
19 MAR 2005
12:34
PRESS
TO EDIT
RANGE
AVG
DIAGS ALARM
The Zero/Span Period Hours screen defines the period or interval between
zero/span checks. Periods between 0 and 999 hours are acceptable. To turn
the zero/span check off, set the period to 0.
In the Main Menu, choose Calibration > Zero/Span Check > Period
Hours.
ZERO/SPAN CHECK:
>NEXT TIME 01JAN06 12:00
PERIOD HR
TOTAL DURATION HR
ZERO DURATION MIN
SPAN DURATION MIN
PURGE DURATION MIN
ZERO/SPAN AVG SEC
ZERO CAL RESET
RANGE
AVG
SPAN CAL RESET
ZERO/SPAN RATIO
ZERO/SPAN PERIOD:
CURRENTLY:
SET TO:
MOVE CURSOR
CHANGE VALUE
RANGE
AVG
DIAGS ALARM
6
1.1
30
30
5
60
OFF
DIAGS ALARM
OFF
1
NEXT DATE AND TIME:
19 MAR 2005
SETTING: DAYS
SET MONTHS
CHANGE VALUE
SAVE VALUE
RANGE
AVG
DIAGS ALARM
024 HRS
025
HRS ?
SAVE
12:34:56