C-Link Protocol Commands
Record Layout Definition

Selection Table

Button Designator
B-46
Model 43i Trace Level-Enhanced Instruction Manual
as a zero-based index into the translation table to determine the string to
display.
Then there appears an optional selection table within parentheses '(...)'.
This is a string of numbers separated by spaces '(0 1)'. The selection table
lists the translation table entries which the user may select from when
setting the parameter. This is not necessarily the same as the entries which
may be displayed.
Then there appears an optional button designator. This will be one of 'B',
'I', 'L', 'T', or 'N'.
B- Indicates a button which pops up an input dialog prompting the
user for a new value using the designated input format. The input
format is specified from the 'B' through the subsequent semicolon.
I—Indicates a button which pops up a selection list with input
translation. That is, the values read are translated before they are
compared to the selection list options.
L—Indicates a button which pops up a selection list without any
translation. The output value is number of the selected option.
T—Indicates a button which pops up a selection list with output
translation. The number of the option selected is used as an index into
the translation table to generate an output string.
N—Indicates a button which only sends the subsequent command to
the instrument. No user-prompting happens.
Some examples ('\n' is the C syntax for an end-of-line character):

Examples

'
Concentrations\n
This is a single text-only line.
'
'
\n
This is a single blank line.
'
NO:3s\n
'
'