Thermo Scientific 43i Instruction Manual page 249

Pulsed fluorescence so2 analyzer

Advertisement

Thermo Fisher Scientific
conv temp
This command reports the current SO
example that follows reports that the current converter temperature is 45
°C.
Send:
conv temp
Receive:
conv temp 45.0 deg C
flow
This command reports the current sample flow. The example that follows
reports that the current sample flow is 0.503 liters/minute.
Send:
flow
Receive:
flow 0.503 1/m
internal temp
This command reports the current internal instrument temperature. The
first reading is the temperature being used in instrument calculations. The
second temperature is the actual temperature being measured. If
temperature compensation is on, then both temperature readings are the
same. If temperature compensation is off, a temperature of 30 °C is used as
the default temperature even though the actual internal temperature is 27.2
°C. The example that follows shows that temperature compensation is on
and that the internal temperature is 27.2 °C.
Send:
internal temp
Receive:
internal temp 27.2 deg C
pmt voltage
This command reports the current PMT voltage. The example that follows
reports that the current PMT voltage is -510 volts.
Send:
pmt voltage
Receive:
pmt voltage -510 volts
pres
This command reports the current reaction chamber pressure. The first
pressure reading is the pressure reading being used in instrument
calculations. The second pressure is the actual pressure reading being
measured. If pressure compensation is on, then both pressure readings are
the same. If pressure compensation is off, a pressure of 760 mmHg is used
as default pressure even though the actual pressure is 753.4 mmHg. The
example that follows shows that the actual reaction chamber pressure is
753.4 mmHg.
C-Link Protocol Commands
converter temperature. The
2
Model 43i Trace Level-Enhanced Instruction Manual