C-Link Protocol Commands
Calibration
B-24
Model 43i Trace Level-Enhanced Instruction Manual
set cal so2 bkg
This command will auto-calibrate the SO
follows shows a successful auto-calibration of the SO
Send:
set cal so2 bkg
Receive:
set cal so2 bkg ok
so2 coef
coef so2
high so2 coef
low so2 coef
These commands report SO
and low range coefficients in dual or auto range mode. If the mode is
incorrect, the instrument responds with "can't, wrong settings". The
example that follows reports that the SO
Send:
so2 coef
Receive:
so2 coef 1.200
set so2 coef value
set coef so2 value
set high so2 coef value
set low so2 coef value
These commands set the SO
where value is a floating-point representation of the coefficient. The
example that follows sets the SO
Send:
set so2 coef 1.200
Receive:
set so2 coef 1.200 ok
so2 bkg
This command reports the current SO
follows reports that the SO
Send:
so2 bkg
Receive:
so2 bkg 21.4 ppb
set so2 bkg value
This command is used to set SO
value, where value is a floating-point representation of the background in
current selected units. The example that follows sets the SO
21.4 ppb.
background. The example that
2
coefficients in single range mode, or the high
2
coefficient is 1.200.
2
coefficients to user-defined values to value,
2
coefficient to 1.200.
2
backgrounds. The example that
2
background is 21.4 ppb.
2
backgrounds to user-defined values to
2
background.
2
background to
2