diag volt iob
This command reports the diagnostic voltage measurements on the I/O
expansion board. The sequence of voltages is: Positive 24, positive 5,
positive 3.3, and negative 3.3. Each voltage value is separated by a space.
Send:
diag volt iob
Receive:
diag volt iob 24.1 4.9 3.2 -3.2
lamp setpoint
This command reports the set lamp setpoint. The example that follows
reports the lamp setpoint is 1000 counts.
Send:
lamp setpoint
Receive:
lamp setpoint 1000 counts
set lamp setpoint value
value = 0 < value < 4095
These commands set the lamp setpoint in counts. The example that follows
sets the lamp setpoint to 100 counts.
Send:
set lamp setpoint 100
Receive:
set lamp setpoint 100 ok
lamp int
This command reports the current flash lamp intensity. The example that
follows reports that the current flash lamp intensity is 3867 Hz.
Send:
lamp int
Receive:
lamp int 3867 Hz
lamp voltage
This command reports the current flash lamp voltage. The example that
follows reports that the current flash lamp voltage is -810 volts.
Send:
lamp voltage
Receive:
lamp voltage -810 V
led status
led
These commands report the status of the optical test LED on or off. The
example that follows reports that the optical test LED is on.
Send:
lamp status
Receive:
lamp status on
C-Link Protocol Commands
Model 43i Trace Level-Enhanced Instruction Manual
Diagnostics
B-15