Input Board Test; Temperature Calibration; Analog Output Calibration - Thermo Scientific 43i Instruction Manual

Pulsed fluorescence so2 analyzer

Advertisement

Input Board Test

Temperature
Calibration
Analog Output
Calibration
Thermo Fisher Scientific
The Input Board Test screen is used to manually adjust the input board
gain. The input board test is visible only when the instrument is in service
mode. For more information on the service mode, see "Service Mode"
earlier in the chapter.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Input Board Test.
GAIN
FREQ = :
RANGE
The Temperature calibration screen allows the user to view and set the
ambient temperature sensor calibration. The temperature calibration is
visible only when the instrument is in service mode. For more information
on the service mode, see "Service Mode" earlier in the chapter.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Temperature Calibration.
CALIBRATE AMBIENT TEMP:
CURRENTLY:
SET TO:
CHANGE VALUE
RANGE
The Analog Output Calibration menu is a selection of analog output to
calibrate, and allows the user to select the calibration action zero or span.
The analog output calibration is visible only when the instrument is in
1
- MANUAL:
0
CHG GAIN
AVG
DIAGS ALARM
30.0
25.0
MOVE CURSOR
AVG
DIAGS ALARM
Model 43i Trace Level-Enhanced Instruction Manual
Operation
Service Menu
?
o
C
o
C?
SAVE
3-71