7 Error messages
7.1 List of error messages
38
7

Error messages

Errors during the self-test are indicated by a five-digit number lighting up.
The background color of the display is red.
CAUTION
If an error re-occurs after the unit has been switched off and switched
on again, please call your service engineer.
Tell the service engineer which error message was displayed.
7.1

List of error messages

Error code
Reason and measures
E3 04 30
Release error - the release button may have been
pressed during switch on
Switch the device OFF and then ON again
If the error persists, call a service engineer and report the
error code
E1 11 88
Display mode ACTIVE - X-ray cannot be release
Call a service engineer and report the error code
E1 04 03
Internal error
E1 04 04
Press any button to acknowledge the error
E1 04 06
If the error persists, call a service engineer and report the
E6 04 02
error code
E5 04 50
Internal error
E6 01 41
Switch the device OFF and then ON again and repeat the
E6 01 61
exposure
E6 01 62
If the error persists, call a service engineer and report the
E6 04 01
error code
E6 04 10
E6 04 11
E6 04 12
E6 04 20
E6 04 21
E6 04 40
E6 04 41
E6 04 42
E7 01 01
E7 01 21
E7 01 51
E5 01 02
Internal error
E5 01 12
Call a service engineer and report the error code
E5 01 14
E5 01 22
E5 01 32
E5 01 42
E6 01 11
E6 01 13
E6 01 23
E6 01 31
Sirona Dental Systems GmbH
Operating Instructions HELIODENT
62 14 717 D3507
D3507.201.01.13.02 08.2015