Replacing The Ceph Sensor Holder - Sirona ORTHOPHOS XG 3Dready Service Manual

Table of Contents

Advertisement

6.19 Replacing the ceph sensor holder

3.
D
C
Removing the defective sensor holder
Installing the new sensor holder
6 – 60
6.19
D
2.
1.
Remove the sensor.
2.
Loosen the three screws A and remove the connection socket B. Cau-
tion: Cable L35!
3.
Detach the cable shield from terminal C and the connector screw connec-
tion (loosen the two screws D from the rear) and thread the connector to
the rear through the connection socket.
Install the new sensor holder in the reverse order of removal.
i
NOTE
Don't forget to reattach the cable shield.
ATTENTION: What to do after replacement?
– Perform the CEPH system adjustment (see section 4.5).

Replacing the ceph sensor holder

B
A
1.
Tab 6
59 38 399 D3352