3.3

PERFORMANCE SPECIFICATIONS

Accuracy and Precision

IMMUNOPREP Reagent System Dispensing

Reagent
IMMUNOPREP A
IMMUNOPREP B
IMMUNOPREP C

Pipetting Module - Reagents

Dispense Volume (µL)
1000
200
100
50
20
10

Pipetting Module - Specimens***

Dispense Volume (µL)
200
100
50
20

Flow-Count Fluorospheres Dispensing

Dispense Volume (µL)
100
* Accuracy specifications are based on the mean weight of water based on 10 replicates
(n = 10).
** Coefficient of variation is based on the mean and standard deviation of 10 replicates
(n = 10).
*** Performance claims for specimen volumes not specified above have not been established.
Note: The performance characteristics stated in the specifications were not established by any
in vitro diagnostic assays. It is the responsibility of each laboratory to verify that the system
works for its applications.
PN 623112BA
Volume (µL)
600
265
100
*Accuracy (g of water)
0.950 - 1.050
0.190 - 0.210
0.095 - 0.105
0.047 - 0.053
0.018 - 0.022
0.009 - 0.012
5
0.003 - 0.008
*Accuracy (g of water)
0.190 - 0.210
0.095 - 0.105
0.047 - 0.053
0.018 - 0.022
*Accuracy (g of water)
0.095 - 0.105
PERFORMANCE SPECIFICATIONS
Precision (CV)
Accuracy (± % of volume)
<5%
<5%
<5%
**Precision (CV)
<5%
<5%
<5%
<5%
<10%
<10%
<25%
**Precision (CV)
<5%
<5%
<5%
<10%
**Precision (CV)
<5%
3
SPECIFICATIONS
±5%
±5%
±5%
3-5