3.3
PERFORMANCE SPECIFICATIONS
Accuracy and Precision
IMMUNOPREP Reagent System Dispensing
Reagent
IMMUNOPREP A
IMMUNOPREP B
IMMUNOPREP C
Pipetting Module - Reagents
Dispense Volume (µL)
1000
200
100
50
20
10
Pipetting Module - Specimens***
Dispense Volume (µL)
200
100
50
20
Flow-Count Fluorospheres Dispensing
Dispense Volume (µL)
100
* Accuracy specifications are based on the mean weight of water based on 10 replicates
(n = 10).
** Coefficient of variation is based on the mean and standard deviation of 10 replicates
(n = 10).
*** Performance claims for specimen volumes not specified above have not been established.
Note: The performance characteristics stated in the specifications were not established by any
in vitro diagnostic assays. It is the responsibility of each laboratory to verify that the system
works for its applications.
PN 623112BA
Volume (µL)
600
265
100
*Accuracy (g of water)
0.950 - 1.050
0.190 - 0.210
0.095 - 0.105
0.047 - 0.053
0.018 - 0.022
0.009 - 0.012
5
0.003 - 0.008
*Accuracy (g of water)
0.190 - 0.210
0.095 - 0.105
0.047 - 0.053
0.018 - 0.022
*Accuracy (g of water)
0.095 - 0.105
PERFORMANCE SPECIFICATIONS
Precision (CV)
Accuracy (± % of volume)
<5%
<5%
<5%
**Precision (CV)
<5%
<5%
<5%
<5%
<10%
<10%
<25%
**Precision (CV)
<5%
<5%
<5%
<10%
**Precision (CV)
<5%
3
SPECIFICATIONS
±5%
±5%
±5%
3-5