Resetting the PM Counts and Changing the Service History

[1] In the service tool, select Monitoring and Control > Imager Monitor.
The Imager Monitor screen displays.
[2] Click:
• Reset Prints To PM
This action also enters the PM record into the service history.
• Reset Prints To Filter Replacement
[3] Perform
"Completing the

Completing the PM

[1] Wait while the processor warms to Ready.
[2] Check the temperature of the drum (see Adjustments and Replacements,
Adjustment").
[3] If necessary, adjust the temperature.
[4] Make a flat-field gray test print.
[5] Compare the test print with the test print made before the PM.
[6] Make a print of a customer image.
[7] Check the quality of the print with the customer.
[8] Leave 2 deodorant filters with the customer.
3-8
PM".
"Processor—Drum Temperature