Konica Minolta DRYPRO SIGMA Service Manual page 203

Medical printer
Hide thumbs Also See for DRYPRO SIGMA:
Table of Contents

Advertisement

Upgrading the Imager Flash Drive with a New Version of System Software – Ghosting
Important
Save the backup file to your laptop computer.
[1] Make a backup of the configuration file for the imager (see
[2] Close the film cartridge.
[3] De-energize the imager.
[4] Remove the back cover (see Adjustments and Replacements,
[5] Install the portable flash drive that contains the ghost software into one of the USB ports on the DRE board.
Note
The USB ports are located behind the lower 2 network connections.
[6] Install the back cover, and then attach the power cord.
[7] Energize the imager to begin ghosting.
The imager beeps 3 times when ghosting has completed.
[8] After ghosting has completed, de-energize the imager.
[9] Remove the back cover, and then remove the portable flash drive.
[10] Install the back cover, and then attach the power cord.
[11] Energize the imager.
[12] Do the microcontrollers need to be updated?
Yes
a. Perform
"Upgrading the Firmware in
the Microcontrollers"
b. Continue with Step 13.
[13] Perform
"Restoring the
9G9002 – 2011-08-15
Continue with Step 13.
Configuration".
Additional Service Procedures
"Making a Configuration
"Back
cover").
USB
Ports (4)
No
Backup").
6-17