Processor Cover

Prerequisites:
T10 Torx driver
Tools:
Removal
Important
The bearings on the ends of each roller are loose. Be sure to remove the correct screws at the side of the processor.
Loosening the wrong screws will cause the bearings or the springs in the retainer to become loose and difficult to
reset.
[1] Remove 2 screws and bushings on each side of the processor cover.
Note
The front retaining rod is held in place by the front screws and is free after you remove them.
[2] Remove the processor cover.
Installation
Reverse the steps in the removal procedure. Make sure that the front retaining rod is secured when you install the
front screws of the cover.
9G9002 – 2011-08-15
• Close the film cartridge. (Touch the Pause button on the display screen.)
• De-energize the laser imager.
• Disconnect the power cord.
• Remove:
Top cover
Right cover
Left cover
Drum Cover Screw, front
Processor Drum
Drum Cover Screw, rear
Adjustments and Replacements
2-23