SERVICE MANUAL
Vacuum Pump
Prerequisites:
T20 Torx driver
Tools:
Removal
Vacuum Pump
Screw
Clamp
[1] Remove the:
(a) Screw from the clamp
(b) Loop clamp from the vacuum pump
Important
When disconnecting the tubing, leave the tie wrap in place.
[2] Disconnect the:
(a) Connector
(b) Tubing
Installation
[1] Connect the:
(a) Tubing
(b) Connector
[2] Place the loop clamp around the vacuum pump.
[3] Insert the screw through the ends of the loop clamp.
[4] Install the screw with clamp and pump onto the frame.
2-32
• Close the film cartridge. (Touch the Pause button on the display screen.)
• De-energize the laser imager.
• Disconnect the power cord.
•
Remove the back
cover.
Connector
Tubing connections
9G9002 – 2011-08-15