SERVICE MANUAL

Film Cartridge Detect Sensor

Remove the DRE motherboard bracket.
Prerequisites:
Tools:
Removal
ESD
Possible damage from electrostatic discharge.
When installing circuit boards, always use a wrist strap to prevent damage from electrostatic
discharge.
[1] Remove the 4 screws near each corner of the film guide, and then remove the film guide.
After you remove the film guide, the 2 film transport guides are visible.
Film Guide
H221_6013HC
Caution
Caution
The imaging roller of the exposure transport can be scratched easily while working in that area. To protect the roller,
cover the roller opening with a sheet of used film.
[2] For better access to the sensor, remove the front and top film transport guides.
(a) Remove the 2 screws at the ends of the front film transport guide, and then remove the guide.
(b) Remove the 4 screws at the corners of the top film transport guide, and then remove the guide.
2-30
• T10 Torx driver
• Small flathead screwdriver (optional)
Front Film
Transport Guide
Top Film
Transport Guide
Film Cartridge Detect
Sensor
Connector terminal