Condition Code: 04-300 – Referenced PLUT is Missing
Web Portal Message
None
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Condition Code: 04-301 – Missing Required Attribute
Web Portal Message
None
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
4-18
Log Message
Referencing Missing
PLUT
The referenced preference look up table (PLUT) specified by tag (2050,0500) in
the film box or image box is not present.
The MIM core software will not process the current print job.
None
False
NORMAL
ERR_MISSING_PLUT
Note
Rationale for not reporting to user: This is a DICOM SCU configuration problem,
not an operational error. The user cannot correct this from the imager. (This
problem likely occurs only on first-time SCU setup.)
Log Message
Required attribute xxxx
(GGGG,EEEE) is
missing from YYYY
SOP (where xxxx
(GGGG,EEEE) is the
tag name and number,
and YYYY is SOP
name)
A required attribute for the specified SOP class is missing.
The named SOP will not be processed. If the SOP is a film session or film box,
the requested page(s) will not print.
None
False
NORMAL
ERR_MISSING_ATTRIBUTE
Note
Rationale for not reporting to user: This is a DICOM SCU configuration problem,
not an operational error. The user cannot correct this from the imager. (This
problem likely occurs only on first-time SCU setup.)
DICOM Printer
StatusInfo
NORMAL
DICOM Printer
StatusInfo
NORMAL
Severity
1-Alert