SERVICE MANUAL
Condition Code: 20-913 – DRE Data Transfer Error Notification
Web Portal Message
None
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
4-44
Log Message
Image transfer failure.
Wrong number of bytes
transferred in the
allotted timeframe.
This notification occurs when the wrong number of pixels is transferred between
the DRE and the DPB in the allotted timeframe. The faulty image prints. The job
fails and the imager attempts to print the image again.
Note
Code 20-915 is generated when this problem is detected on the third successive
film.
None. The system will function normally. [SI-0]
None
True
NORMAL
ERR_DATA_TRANSFR_RETRY