Konica Minolta DRYPRO SIGMA Service Manual page 93

Medical printer
Hide thumbs Also See for DRYPRO SIGMA:
Table of Contents

Advertisement

Condition Code: 04-100 – Limit on Number of Simultaneous Associations
Web Portal Message
None
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Condition Code: 04-101 – The Called AE Title Does Not Correspond to a Recognized Printer
Web Portal Message
None
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
9G9002 – 2011-08-15
Log Message
Maximum Number of
Associations Exceeded
The maximum number of associations configured has been exceeded. The
association is rejected.
The Service Class User (SCU) will not be able to open an association, and
therefore will not be able to deliver a print job.
None
False
NORMAL
ERR_EXCEED_ASSOC
Note
Rationale for not reporting to user: This problem is likely transient. If it is not, the
user cannot correct this from the imager.
Additional DICOM information: There is a specific result/reason code available
for this circumstance: -Rejected transient/3-Local limit exceeded.
Log Message
No Destination
Matching Called AE
Title
The AE title did not map to a destination that is currently available at this
Service Class Provider (SCP).
The SCU will not be able to open an association, and therefore will not be able
to deliver a print job.
None
False
NORMAL
ERR_AE_NOT_RECOGNIZED
Note
Rationale for not reporting to user: This is a DICOM SCU configuration problem,
not an operational error. The user cannot correct this from the imager.
Additional DICOM information: There is a specific result/reason code available
for this circumstance: 1-Rejected permanent/7-Called AE title not recognized.
DICOM Printer
StatusInfo
NORMAL
DICOM Printer
StatusInfo
NORMAL
Diagnostics
Severity
1-Alert
Severity
1-Alert
4-15

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents