Konica Minolta DRYPRO SIGMA Service Manual page 144

Medical printer
Hide thumbs Also See for DRYPRO SIGMA:
Table of Contents

Advertisement

SERVICE MANUAL
Condition Code: 27-609 – Optics Maximum Laser Power Approaching Limit
Web Portal Message
None
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Condition Code: 27-611 – Optics Laser Gain Adjustment Failed
Web Portal Message
Optics: Internal
Hardware Failure
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Logged for RMS
4-66
Log Message
Optics calibration
detected Optics
Maximum Power
approaching limit
This notification occurs when the maximum laser power measured by the optics
calibration and diagnostics is approaching the predefined valid limit.
None. The system will function normally. [SI-0]
None
False
NORMAL
ERR_OPTICS_MAX_LASER_POWER_NEAR_LIMIT
Log Message
Optics calibration laser
power gain adjustment
failed
The laser power gain is measured using a voltage on the laser driver. This error
occurs when the laser driver voltage cannot be adjusted to within limits by the
optics calibration after 20 successive attempts.
The imager can only print jobs that were in process when the error occurred.
The spooler can still accept jobs, but the jobs will not print until the error is
resolved. All jobs that were in process when the error occurred will reprint
completely when the error is resolved. [SI-1]
1. Shut down the laser imager, and then restart the imager.
2. If the error persists, call for service.
True
FAILURE
ERR_OPTICS_LASER_GAIN_FATAL_ERROR
[TriggerValues=x][TriggerRange=y,z][TriggerNotes=Laser Voltage: value,
limits]
DICOM Printer
StatusInfo
NORMAL
DICOM Printer
StatusInfo
ELEC DOWN
Severity
3-Notify
Severity
0-Fatal