SERVICE MANUAL
Condition Code: 28-510 – Processor RTD Open Circuit
Web Portal Message
Processor: Internal
Hardware Failure
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Logged for RMS
4-70
Log Message
Processor detected a
sudden change to
infinite resistance in the
RTD circuitry
This error likely occurred in conjunction with the overheat error. The most likely
cause is an opening in the RTD circuit (RTD fractured, drum slip ring failure,
wiring harness disconnected, etc.).
The error clears only when imager power is cycled.
The imager can only print jobs that were in process when the error occurred.
The spooler can still accept jobs, but the jobs will not print until the error is
resolved. All jobs that were in process when the error occurred will reprint
completely when the error is resolved. [SI-1]
1. Shut down the laser imager, then restart the imager.
2. If the error persists, call for service.
True
FAILURE
ERR_PROCESSOR_RTD_OPEN
Logged for RMS: [TriggerValues=a, b][TriggerNotes=RTD zone number,
temperature in degrees C]
[TriggerValues=a, b][TriggerNotes=RTD zone number, temperature in
degrees C]
DICOM Printer
StatusInfo
PROC DOWN
Severity
0-Fatal
9G9002 – 2011-08-15