Checking Image Quality
[1] Ask the customer to send images from the modalities to the laser imager.
[2] Check for image skew. If necessary, adjust the optics assembly (see
Adjustment").
[3] Is the image quality acceptable to the customer?
Yes
Advance to
"Editing the Service
History
Log".
[4] Click the Image Optimization tab on the SCP Configuration screen.
[5] At the top of the screen, select an SCU from the drop-down list.
Note
• The information for each entry in the SCU drop-down list includes, in order: IP Address and AE Title of the SCU
that will send the image, and AE Title of the destination.
• The parameter values on the screen apply to the modality identified in the SCU list.
• Perform Image Optimization and Other Options for each SCU in the drop-down list.
[6] Click Modify.
[7] To change a parameter, type the new value or select it from the drop-down list.
Note
• The value selected will be the default for this SCU.
• The value in the list to the right of the parameter value indicates whether the value sent from the SCU is
overridden.
– If you select Always Apply, the parameter on the screen overrides the parameter the SCU sends.
– If you select No, the SCU parameter is applied. Normally, you should select No to allow the SCU to override
the default selected.
9G9002 – 2011-08-15
No
Continue with Step 4.