Setting the Laser Imager to be a Destination
[1] Select Configuration > DICOM SCP > Destination.
Important
The AE Title is the name by which the modalities recognize the laser imager.
[2] Check with the customer to determine the AE Title to use.
[3] Click Modify.
[4] Type the selected AE Title in the window.
[5] Click the Media tab.
[6] In the Available column, select the check box next to the Media Type that will be used with the imager.
Note
Under normal operations, all media types should be allowed.