SERVICE MANUAL

Configuring the Laser Imager

Important
You must restart the laser imager from the service tool to permanently save configuration information.
Setting the Language
Important
The language is set by the customer in the Web Portal.
Checking the Values for the DICOM Source Communications
[1] Connect the laptop computer to the laser imager (see Installation Instructions,
the Laser
Imager").
[2] Select Configuration > DICOM SCP > Communications.
[3] Check:
• Port Number is 5040.
• Association Limit is 2.
[4] Are the values correct?
Continue with
"Setting the Laser Imager to be a
Destination".
6-2
Yes
No
a. Click Modify.
b. Type the correct values.
c. Click Save.
d. Continue with
"Setting the Laser Imager to
be a
Destination".
"Connecting Your Computer to
9G9002 – 2011-08-15