Konica Minolta DRYPRO SIGMA Service Manual page 54

Medical printer
Hide thumbs Also See for DRYPRO SIGMA:
Table of Contents

Advertisement

SERVICE MANUAL
[3] Remove the left and right side rollback springs.
[4] Move the rollback assembly to the fully open position.
[5] From the left side of the imager, remove the E-ring from the end of the rollback knob, and then remove the knob.
This allows more clearance to remove the pickup assembly.
[6] Remove 2 screws each on the left and right side of the imager.
Pickup Assembly
Mounting Screws
(2 of 4)
[7] Pull the film pickup assembly through the slot on the left side of the laser imager and place it on a flat, stable
surface.
Installation
Reverse the steps in the removal procedure.
2-34
Rollback Springs
Pickup Assembly
E-ring
Manual
Rollback Knob
9G9002 – 2011-08-15