Restoring the Configuration
[1] Log on to the service tool (see
[2] Select Utilities > Backup/Restore.
[3] Click the Restore tab.
[4] Select the type of information to be restored from the Restore Level drop-down list.
[5] Choose how you want to restore the backup data:
From flash drive
Click Restore.
Important
You must restart the imager from the service tool.
[6] Select Utilities > Session > Restart.
[7] Click Restart.
Wait for the imager to restart.
[8] If you are finished with the service tool:
(a) Disconnect the laptop computer from the imager.
(b) Install the top cover.
9G9002 – 2011-08-15
"Logging On to the Service
Upload stored configuration data from laptop
a. Select the Upload Configuration File check box.
b. Click Select.
c. In the File Upload window, click Browse.
d. When the Choose File window displays, select the .bin file.
e. Click Open, and then click Upload.
Additional Service Procedures
Tool").