To Adjust:
[1] Log on to the service tool (see Additional Service Procedures,
[2] Select Configuration > Hardware > Processor Configuration.
[3] Click Modify.
[4] In the Temperature field, enter the temperature measured by the meter.
Note
The Offset field indicates the offset value between the Set Point field and the Temperature field.
[5] Click Set Temperature.
[6] Allow approximately 10 minutes for the temperature of the processor drum to become stable.
[7] Use the meter to check the temperature of the processor drum. The temperature must be 123.5–124.5 °C.
[8] When the temperature is between 123.5–124.5 °C, close the Processor Configuration window.
[9] Log off from the service tool (see Additional Service Procedures,
[10] Install the back cover.