Contents of the Log

Each line of the log documents an event. Each event has 18 columns of information. Only five of the
columns, listed below, are useful for diagnostics by field service personnel. The other data is for use by
Software Engineering.
Column Title
Event Type
Time
Code
Text
App
9G9002 – 2011-08-15
Provides the same information in the Level column—the level provided for the
event. The levels from highest to lowest are: Trace (Level 6), Debug (5), Diag
(4), Info (3), Warning (2), Error (1), and Fatal Error (0).
Provides the time of the event, down to 1/100th of a second. The time is used
to locate events in the log.
The code for an event, normally an error code
Describes the event
The software application in which the event occurred
Function
Diagnostics