SERVICE MANUAL
[11] Take a sheet of the largest size film available and cut it lengthwise to make a strip 8 cm (3 in.) wide.
Note
The strip is used to cover and protect the imaging assembly from debris during cleaning.
[12] Fold the film strip lengthwise to form a V.
[13] From the left side of the imager, insert the strip over the imaging roller to cover it.
Clean the processor
[14]
• Wipe the processor drum with alcohol wipes in one direction with an even steady pressure. Rotate the
processor drum while wiping.
• Check the processor drum for imperfections, gouges, and areas of excessive buildup of debris.
Clean and inspect the processor cover
[15]
[16] Pivot the processor drum cover closed.
Pivot the heat shield
[17]
[18] Remove the felt pad assembly from the diverter assembly (see Adjustments and Replacements,
Assembly").