SERVICE MANUAL
Condition Code: 27-601 – Invalid Power Monitor Offset
Web Portal Message
Optics: Calibration
Failed
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Logged for RMS
Condition Code: 27-604 – Unable to Adjust Laser Dynamic Range
Web Portal Message
Optics: Calibration
Failed
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Logged for RMS
4-64
Log Message
Optics Calibration
Failed. Power Monitor
Offset outside valid
range.
This error occurs when the power monitor offset measured by the optics
calibration and diagnostics is outside the predefined valid range.
The error clears when a calibration is requested or power is cycled.
The imager can only print jobs that were in process when the error occurred.
The spooler can still accept jobs, but the jobs will not print until the error is
resolved. All jobs that were in process when the error occurred will reprint
completely when the error is resolved. [SI-1]
1. Shut down the laser imager, and then restart the imager.
2. If the error persists, call for service.
True
FAILURE
ERR_INVALID_POWER_MONITOR _OFFSET
[TriggerValues=x][TriggerRange=y,z][TriggerNotes=Power Monitor Offset:
value, limits]
Log Message
Optics calibration failed
to Adjust Laser
Dynamic Range within
limits
This error occurs when the laser dynamic range measured by the optics
calibration and diagnostics is outside the predefined valid range after 5
successive adjustment attempts.
The error clears when a calibration is requested or power is cycled.
The imager can only print jobs that were in process when the error occurred.
The spooler can still accept jobs, but the jobs will not print until the error is
resolved. All jobs that were in process when the error occurred will reprint
completely when the error is resolved. [SI-1]
1. Shut down the laser imager, then restart the laser imager.
2. If the error persists, call for service.
True
FAILURE
ERR_CANT_ADJUST_LASER_DYNAMIC_RANGE
[TriggerValues=x][TriggerRange=y,z][TriggerNotes=Laser Dynamic Range:
value, limits]
DICOM Printer
StatusInfo
ELEC DOWN
DICOM Printer
StatusInfo
ELEC DOWN
Severity
0-Fatal
Severity
0-Fatal
9G9002 – 2011-08-15

Hide quick links: