Film Supply Status Messages

The following table describes how the display screen and Web Portal indicate film supply status to the
user:
Film Supply
State
Failed
Calibrating
Film cartridge
empty
Manual Mode
Invalid film
cartridge
Ready
Requires
Calibration
Film cartridge
not detected
4-6
Display Screen
Web Portal
Display
Failed
An error occurred that affects normal
operation. The film cartridge is currently
not usable. Reinsert the cartridge. If the
error recurs, insert a new film cartridge.
Calibrating
A calibration is in progress for this film
cartridge.
Empty and/or
The film cartridge is inserted but the
sheet count of 0
sheet count is 0. Insert a new film
cartridge.
AIQC Off (with
The film in this cartridge does not meet
normal tray
AIQC standards. However, the laser
information)
imager prints if ready.
Invalid Film Tray
There is a film cartridge in the film
supply, but it does not contain a liner/
RF tag. Install a new film cartridge.
Normal Tray Info
The film cartridge is ready to use.
Requires
The film cartridge must be calibrated
Calibration
before the laser imager can print.
Initiation of the calibration may require
user action.
No Film Tray
The film cartridge is not fully inserted
into the laser imager. Insert the
cartridge.
Description
9G9002 – 2011-08-15