[5] Is the Date correct?
Advance to Step 15.
[6] Select System Information > User Information.
[7] Call your Technical Support Center to gain temporary Level 5 access.
[8] Click Modify.
[9] In the Authorization Key field, type the key from the Technical Support Center.
[10] Click Save.
Note
You will have Level 5 access for 24 hours.
[11] Return to the Clock Configuration screen.
[12] Click Modify, and then set the correct date.
[13] If daylight savings time applies in the area of installation, select the Automatically adjust clock for daylight
savings time check box.
[14] Click Save.
[15] Advance to
"Setting the Host Name and the IP
9G9002 – 2011-08-15
Yes
Continue with Step 6.
No
Address".