SERVICE MANUAL
Condition Code: 21-177 – Film Cartridge Failed to Close
Web Portal Message
Film Cartridge Failed to Close
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Logged for RMS
4-54
Log Message
Rollback Home Sensor
(S2U) failed to activate
in specified time
This error occurs only when attempting to close a film cartridge. (Note that an
open cartridge operation always closes the cartridge first, before attempting to
open the cartridge.) This error occurs for all rollback close failure conditions.
Rollback home sensor (S2x) failure is the most likely cause, but the error can
also occur from rollback motor (M6x) failure.
The error is cleared when the film supply is closed (as indicated by the supply
drawer interlock switch i1x), or when power is cycled.
Cannot print from the loaded film cartridge. The spooler is still able to accept
and queue jobs. [SI-2].
1. Open the film supply.
2. Remove the left cover,
3. Use the blue knob to rotate the cover roller to the home position.
4. Manually close the cartridge to prevent film fogging.
5. Close the film supply.
6. If the error persists, call for service.