Condition Code: 21-178 – Rollback Failed to Leave Home
Web Portal Message
Rollback Failed to
Leave Home
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Logged for RMS
Condition Code: 21-624 – Bad Densitometer Data
Web Portal Message
Film Supply: Film
Calibration Failure
Description
System Impact
User Action
Show at DS
DICOM Printer
Status
Unique Symb. ID
Logged for RMS
9G9002 – 2011-08-15
Log Message
Rollback Home Sensor
(S6U) failed to deactivate
in specified time
This error occurs only when attempting to open a film cartridge. This error
occurs if the rollback mechanism movement from its home position toward open
is not confirmed by the rollback home sensor (S2x). Because no rollback
movement was detected, either the rollback motor (M6x) failed or the sensor
failed. The error is cleared when the film supply is closed (as indicated by the
supply drawer interlock switch i1x), or when power is cycled.
Cannot print from the loaded film cartridge. The spooler is still able to accept
and queue jobs. [SI-2].
1. Insert a new film cartridge.
2. If the error persists, call for service.
True
WARNING
ERR_UP_ROLLBACK_NOT_LEAVE
[CartridgeInfo=bbeeee-cjj-d-iii][FilmSize=size]
Log Message
Film Supply:
Calibration Failed
This error occurs when the density readings that come from the calibration print
are not monotonic or the densitometer did not find exactly 21 steps in the
calibration wedge. The most likely cause is fogged or expired film. The error is
cleared when a film cartridge is removed and then re-inserted.
Cannot print from the loaded film cartridge. The spooler is still able to accept
and queue jobs. [SI-2]
1. Insert a different film cartridge.
2. If the error repeats, restart the imager.
3. If the error persists, call for service.
True
WARNING
ERR_UP_BAD_DENSITOMETER_DATA
[CartridgeInfo=bbeeee-cjj-d-iii][FilmSize=size][TriggerValues=a,b,c,...,u]
[TriggerNotes=Optical Density values, from max to min]
DICOM Printer
StatusInfo
BAD SUPPLY MGZ
DICOM Printer
StatusInfo
CALIBRATION ERR
Diagnostics
Severity
1-Alert
Severity
1-Alert
4-55

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents